ProStar325 Service Manual
84
Publication date 8 May 2003
read at that wavelength).
The results from wavelength#2 optional collections are available
in the status message as well as the method run data stream.
Autozeroing can be requested at both wavelengths, however,
peak sense is not valid in Dual Wavelength mode.
Bunching of data is also not available in dual wavelength
collection.
4.4
Initialization Sequence
The detector goes through the following routine when it is
powered up;
Initialize all ports, interrupts & PCI bus
Test on-board SDRAM
Self check CRC of boot block
Check CRC of flash memory main code
Clear relay outputs and SYNC signal outputs
Check the position of the diagnostic rotary switch and perform
the test
Perform normal start up checks
Perform power supply check
Activate READY_OUT SYNC signal
4.5
LEDs and diagnostic display
The Processor PWB is fitted with various LEDs to indicate
functions and provide diagnostic information to CSRs.
Under normal circumstances four green LEDs should be lit -
LED5
+5V
OK
LED6
+9V
OK
LED7
-9V
OK
LED8
+24V
OK
CPU reset status indicated by a red LED
LED1
Reset line is active (processor halted)
Ethernet communications is indicated by a green LED
LED4
Flashing during communications
4.5.1
Seven segment displays
There are two seven segment displays LED 2 & 3 which provide
the status of firmware and fault indication which varies
according to the instrument status. During initialisation, the
following numbers will appear on the two seven-segment
displays;
StockCheck.com
Summary of Contents for ProStar 325 UV-Vis
Page 9: ...ProStar325 Service Manual Publication date 8 May 2003 9 S t o c k C h e c k c o m ...
Page 19: ...ProStar325 Service Manual Publication date 8 May 2003 19 S t o c k C h e c k c o m ...
Page 34: ...ProStar325 Service Manual 34 Publication date 8 May 2003 S t o c k C h e c k c o m ...
Page 71: ...ProStar325 Service Manual Publication date 8 May 2003 71 S t o c k C h e c k c o m ...
Page 72: ...ProStar325 Service Manual 72 Publication date 8 May 2003 S t o c k C h e c k c o m ...
Page 118: ...ProStar325 Service Manual 118 Publication date 8 May 2003 S t o c k C h e c k c o m ...
Page 149: ...ProStar325 Service Manual Publication date 8 May 2003 149 S t o c k C h e c k c o m ...