LED
3
LED
2
0
0
0
0
1
1
1
0
2
4
5
0
1
3
4.6
Diagnostic select switch
A 16-position thumbwheel switch is used to select service
diagnostic routines.
#
DESCRIPTION
0
NORMAL INSTRUMENT OPERATION
1
FLASH MEMORY CRC/ CHECKSUM BOOT BLOCK
Continuously checks the boot block program.
Display ‘10’ if OK
Display ‘11’ if ERROR
2
RAM INTEGRITY test
Continuously loops through all SDRAM
Display ‘20’ if OK
Display ‘22’ if ERROR
3 ADC/DAC
test
Continuously checks the main ADC operation
4
ACCESSORY OUTPUT test
Checks all SYNC and relay outputs by toggling in a two second period.
Display ‘40’ during this test
5
Not used at present
6
MONOCHROMATOR MOTOR test
Initializes the stepper motor and sets the monochromator to somewhere near
theoretical zero order. It then continuously cycles at full speed back and forth of this
position with a 1second interval between direction changes.
Display ‘66’ while the motor is resetting
Display ‘60’ while it is being cycled
StockCheck.com
Summary of Contents for ProStar 325 UV-Vis
Page 9: ...ProStar325 Service Manual Publication date 8 May 2003 9 S t o c k C h e c k c o m ...
Page 19: ...ProStar325 Service Manual Publication date 8 May 2003 19 S t o c k C h e c k c o m ...
Page 34: ...ProStar325 Service Manual 34 Publication date 8 May 2003 S t o c k C h e c k c o m ...
Page 71: ...ProStar325 Service Manual Publication date 8 May 2003 71 S t o c k C h e c k c o m ...
Page 72: ...ProStar325 Service Manual 72 Publication date 8 May 2003 S t o c k C h e c k c o m ...
Page 118: ...ProStar325 Service Manual 118 Publication date 8 May 2003 S t o c k C h e c k c o m ...
Page 149: ...ProStar325 Service Manual Publication date 8 May 2003 149 S t o c k C h e c k c o m ...