ProStar325 Service Manual
Publication date 8 May 2003
97
procedures are necessary. Removal and replacement procedures
are described in this section of the manual.
5.2.2
Visible lamp assembly
When this option is fitted, the visible lamp mounts on top of the
optics module and is held in place by two screw terminal
connections. These are in a fixed position therefore allowing
replacement of the lamp without the need for alignment. Removal
and replacement procedures for this lamp are described in this
section.
5.2.3
Monochromator
The monochromator is a sealed unit that supplies a single
wavelength beam of light directed through the flowcell and
conforms within 1nm to a specifically selected wavelength.
The entrance mask produces a defined beam of polychromatic UV
and visible light before it gets collimated by the first collimating
mirror. The beam is then directed onto a holographic grating that
disperses it into a continuous spectrum. A stepper motor provides
direct drive with micro stepping to rotate the grating and obtain
the desired wavelength. The beam is then directed onto the exit
collimating mirror before passing through the exit mask and onto
the detectors. The exit mask produces a very narrow bandwidth
of light.
Exit mask
Entrance
mask
Monochromator
section
StockCheck.com
Summary of Contents for ProStar 325 UV-Vis
Page 9: ...ProStar325 Service Manual Publication date 8 May 2003 9 S t o c k C h e c k c o m ...
Page 19: ...ProStar325 Service Manual Publication date 8 May 2003 19 S t o c k C h e c k c o m ...
Page 34: ...ProStar325 Service Manual 34 Publication date 8 May 2003 S t o c k C h e c k c o m ...
Page 71: ...ProStar325 Service Manual Publication date 8 May 2003 71 S t o c k C h e c k c o m ...
Page 72: ...ProStar325 Service Manual 72 Publication date 8 May 2003 S t o c k C h e c k c o m ...
Page 118: ...ProStar325 Service Manual 118 Publication date 8 May 2003 S t o c k C h e c k c o m ...
Page 149: ...ProStar325 Service Manual Publication date 8 May 2003 149 S t o c k C h e c k c o m ...