ProStar 335 Photodiode Array Detector
109
Appendices
Specifications
General specifications
Diagnostics:
Built-in self-test and diagnostics
Time programming:
Time-programmable wavelength, recorder attenuation,
auto-zero, response time, peak sense, time slice, pulse and
external event relays. Response time is programmable at
time=zero only.
Wavelength:
UV (Deuterium) lamp and visible (quartz iodide) lamp, 190–
950 nm
Flowcell:
Optional 9 x 0; 9 x 1; 4 x 0; 4 x 0.15
Pressure:
1000 psi maximum on all flowcells
Response time:
0.05, 0.5, 1.0, and 2.0 sec
Spectral bandwidth:
1, 2, 4, 8 or 16 nm
Recorder output:
1 V FS
Data system output:
1 V FS
Peak Sensing:
Peak start, Peak end, Valley, Time Slice and Pulse will activate
the Peak Sense relay and generate event marks on the
recorder chart.
External
communication:
To other instruments by synchronization cables.
Contact closure
outputs:
Four time-programmable external event relays, one Peak
Sense relay, three synchronization signals: READY OUT,
START OUT and FAULT OUT