Before connecting power to the detector, ensure that the power
switch on the front of the instrument is OFF (the rocker switch
{
is pressed). The power switch is a rocker switch that connects
from the front of the detector directly to the power receptacle on
the rear panel. Plug one end of the power cord into the power
receptacle on the rear panel and the other end into your AC
power source.
WARNING
It is important for operator safety and proper operation of the ProStar 335 detector that
you use the power cord supplied, together with a properly grounded 3-wire power strip.
All devices in the liquid chromatography system should be
connected to the same power source, using a properly grounded
(third wire to earth) multiple outlet power strip. Power strips for
operation at 125 VAC, 50-60 Hz are available from Varian, part
numbers R0-38602C-B6 (5 outlets) or R0-03824C-B6 (6 outlets).
Fuses
T3.15AH250V, IEC 127 sheet 5, 5 x 20 mm
NOTE:
For safety reasons, any other internal fuse is not operator accessible, and
should be replaced only by Varian-authorized personnel.
Mains inlet coupler
3/2A 120/250 VAC 50–60 Hz IEC type
Mains power cord
Australia
10 A 250 VAC
Complies with AS3112
USA
10 A 125 VAC
Complies with NEMA 5–15P
Europe
6A 250 VAC
Complies with CEE7 sheet vii or
NFC61.303 VA NFC61
NOTE:
Do not turn on the AC power yet. All required external devices and
hydraulics must first be connected.