I
NSTALLATION
ProStar 335 Photodiode Array Detector
55
The Peak relay is software programmable for duration, delay and
active sense. At power up, the relay contact will be set to the
inactive state (as defined by the value of the ‘active sense’
parameter stored in the detector). Upon being triggered, relay
activation will occur for the time interval equal to the ‘Peak Sense
duration’ parameter as stored in the method.
The Peak relay can be activated from any of the following
sources (only one source can be active at any one time):
Time Slice event – Once time slice has been turned on, it
will provide a periodic activation of the Peak Sense relay
at an interval defined in ‘Time Slice period’ within the
method. Time Slice can be turned on and off by time.
Pulse event – A single timed programmed activation of
the Peak Sense relay as defined in the method.
To connect open-ended wires to the relay signals available at J4,
use the optional Relay Interface board (part number
02-101875-90, refer to Figure 6). Simply plug this PWB into the
rear panel connector.