I
NSTALLATION
ProStar 335 Photodiode Array Detector
63
The ProStar 335 detector is provided with its IP address by the
Star or Galaxie Workstation software running on your network.
The software used to send IP addresses to TCP/IP devices is
called a BOOTP Server (Bootstrap Protocol Server). You will see
the term BOOTP Server mentioned in this section. If you are
connecting your ProStar 335 and Workstation to a company
network, your Network Administrator should be made aware that
the ProStar 335 will require a BOOTP Server for IP address
assignment.
Where to begin
Before beginning the configuration process, you should know
whether you will be attaching your Workstation and ProStar 335
detector to a company network (as described in the previous
section) or an isolated network only used for instrument control. If
you have a Network Administrator on site, you may wish to ask
which configuration is recommended.
If an Ethernet card has not been installed and configured on your
PC, read ‘Installing and Configuring the Ethernet Card in Your
PC‘ (next section) and one of the following:
No Company Network: Configuring TCP/IP Parameters
with No Company Network
Company Network: Configuring TCP/IP Parameters for a
Company Network