S
YSTEM
D
ESCRIPTION
ProStar 340 UV/Vis Detector
9
10. Pressing the top half of this switch to rock it upward to “ON”
sets the full scale recorder output to 10 mV when Switches 8
and 9 are “OFF”.
11. Filter By-Pass Switch: A two-position rocker switch controls
a bypass circuit for the Second Order Bessel filter. Pressing
the top half of this switch so that it is “ON” disables the front
panel Rise Time Control and results in an effective rise time
of 0.1 seconds.
12. Integrator Offset Switch: A two-position rocker switch
provides an addi10 mV offset to the fixed 1 AU/V
signal of the integrator output when the switch is in the ON
position. This switch does not affect the recorder output.
This integrator offset is supplied to aid integrators incapable
of zeroing for a negative drifting baseline. In most cases, it
need not be used.
13. Recorder Offset Switch: A two-position rocker switch
provides a +10% fixed offset to the recorder output. This
offset is independent of full scale range or full scale voltage.
This recorder offset does not affect the integrator output.
14. Power Switch: A two-position rocker switch turns the
instrument on and off. Pressing the top half of this switch to
rock it upward powers up the unit. Pressing the bottom half
of the switch to rock it downward shuts the unit off.
15. Power Connector: A three-pin receptacle is provided that
accepts a standard modular power line cord.
16. Fuse Block: Pry out this block to allow access to the fuses
and voltage control. It contains one 1.0 Amp slow-blow fuse
(for 100-120 Vac operation) and two 0.5 Amp slow-blow
fuses (220-240 Vac operation).
17. Voltage Control Selector: A four position voltage selector
allows the instrument to be operated at 100, 120, 220, or
240 Vac (50/60 Hz).