S
YSTEM
D
ESCRIPTION
ProStar 340 UV/Vis Detector
13
OPEN
FILTER BYPASS
INT. OFFSET
REC. OFFSET
1V
100mV
10mV
RECORDER
F/S
1 2 3 4 5 6
Figure 7 Recorder Full-scale Voltage Switches
NOTES:
If any of these switches are switched on together, the recorder output
shows uncalibrated full scale. Ensure that only one switch is on at a time.
These switches do not affect integrator output.
Recorder Connections
The recorder cables are connected at the I/O terminal on the
back panel. The cables should have about 1/4" of bare wire or a
spade connector. Connect the positive input of the recorder to
the positive screw on the recorder output. Connect the input of
the recorder to the negative screw on the recorder output.
1
2
3
4
5
6
7
+ – + – + – + – + –
INT
1AU/V
AUTO
ZERO
EVENT
LAMP
RECORDER
OUTPUT
REMOTE
Figure 8 Input/Output Terminal
Integrator Connections
The detector also provides a fixed-span output (1V/AU) for use
with an integrator or data acquisition system. This is independent
of the rise time control, full scale range and voltage settings of
the unit. However, its output will reflect the zeroing of the
autozero circuit. Thus, a changing baseline can be corrected by
pushing the autozero switch.
Connect the input line of the integrator in an identical manner as
was outlined for the strip chart recorder.