14
03-914783-00:3
Power Cord
Power to the detector is provided by a standard modular power
cord assembly. Connect the power cord to the receptacle next to
the fuse block.
Remote Connections
The detector provides external remote control connections. (You
do not need to connect these in order to finish the initial
installation and check-out. In that case, go to the Fluid
Connections section.) Refer to Figure 8 for Remote connections.
Autozero
The autozero connection duplicates the function for the autozero
switch on the Front Panel whenever it is connected to a
momentary contact closure or TTL low.
Connect the triggering device so that the positive (+) line is
connected to the positive pole and the negative (–) line is
connected to the negative pole of the remote Autozero terminal.
The remote autozero is triggered by shunting across its two input
terminals.
Event
The Event connection duplicates the function of the Event Button
on the front panel whenever it is connected to a momentary
contact closure or TTL low.
Connect the triggering device so that the positive (+) line is
connected to the positive pole and the negative (–) line is
connected to the negative pole of the remote Event terminal. The
remote Event is triggered by shunting across its two input
terminals.
NOTE:
If you notice excessive noise when a remote triggering device is
connected or used, there may be a ground loop in the circuit. Ensure that
the remote device has a negative output which is isolated from the earth
ground.