Installation
ProStar 363 Fluorescence Detector
5
Power capacity:
360 VA for the detector. The detector is
used as part of an HPLC system,
therefore the power to be provided to the
system is dependent on the configuration
of the system. Ensure that the power
supply is sufficient to meet the needs of
the entire system.
Ground Terminal
WA R N IN G :
SH O C K H A ZA R D
Ground Properly to Prevent Electric Shock Hazard!
•
Be sure to use the power cable supplied with the
instrument. Use of a different power cable may
result in an electric shock hazard.
•
This instrument is classified as "1" in IEC1010-1
Annex H and "plug-connected type" in IEC1010-1, so
connect the power cable to a grounded 3-wire outlet.
•
If a grounded 3-wire outlet is not available be sure to
provide proper grounding connection.
Provide grounding connection having resistance of less than 100
ohms within a distance of three meters.
Figure 2 Power Receptacle Polarity
Hot
(Line)
Ground
Pin
Neutral
(light blue or white)
Ground (green/yellow or green)
Hot (brown or black)
Hot (Line)
Neutral
15 Amps
120 Vac Nominal
60 Hz
220 Vac Nominal
50 Hz
Ground
Neutral