If your instrument fails to power up when the power switch is
turned on, check that the power cord is properly connected, and
check for power at the wall receptacle.
Location
The ProStar 363 Fluorescence Detector is a sensitive instrument
and should always be handled with the degree of care
appropriate to laboratory instrumentation. After unpacking, place
the detector conveniently near your HPLC system. For normal
operation, the detector should be located on a clean flat surface
away from:
•
heat sources (such as direct sunlight or a heater vent)
•
drafts (such as an open doorway, window, or air-conditioner
vent)
•
corrosive or dusty atmosphere
•
vibration
•
potential liquid spills
Place the detector feet on a firm flat surface. Allow at least four
inches of clear space at the back of the instrument so that the
cooling fan intake is not impeded. The detector should be
installed in a location with the following dimensions:
Minimum width = 450 mm
Minimum depth = 650 mm
The bench that is used for the instrument should be capable of
supporting at least 50 kg.