Installation
ProStar 363 Fluorescence Detector
11
If the ProStar 363 is to send data to a Star Workstation, the
Processor or Data System output must be connected either to a
CIM box, CIM in a pump, or an ADC board. Connect the ProStar
363 to the ADC board with the Fluorescence detector interface
cable (03-936156-01). Connect the ProStar 363 to the CIM input
with the CIM to ProStar 363 cable (03-935922-01).
If you are going to connect the ProStar 363 to some other
instrumentation, you can use the connections on the rear panel
for synchronizing the module with any other HPLC system. The
inputs and output are explained in the table below.
Table 1 Input/Output Terminals
Designation Input/
Output
Function Description
Data
System/
Processor
Output
Analog output for
processor, 1 volt full
scale
The following signals can be output
according to the setting.
•
Fluorescence intensity at each time
point
•
Stored spectral data
Recorder
Output
Analog output for
recorder, 10 mV full
scale
The following signals can be output
according to the setting.
•
Fluorescence intensity at each time
point
•
Stored spectral data
Ready/ Busy
Out
Output
Level change
Level goes high when detector is ready.
Error Out
Not used at this time.
Start In
Input
Contact closure
A contact closure here will start the program
on the ProStar 363.
AutoZero In
Input
The auto zero function
is turned on by the
contact signal input.
The auto zero function is started when the
terminals are shorted for at least one
second.
Marker In
Input
The marker function is
turned on by the
contact signal input.
The marker function is started when the
terminals are shorted for at least one
second. A positive pulse is added to
Recorder Output Signal.
Lamp Off In
Input
The xenon lamp is
turned off by the
contact signal input.
The lamp is turned off when the terminals
are shorted for at least one second. An
alarm sounds and an alert message is
displayed. Pressing clear will stop alarm
and re-light the lamp.