Installation
ProStar 363 Fluorescence Detector
13
Note the arrow stamped on the terminator body. This arrow must
point away from the flow cell outlet port, and toward the waste
receiver. The threaded plastic fittings should be finger-tightened
only enough to prevent leaks.
About 48" of 1/16" Teflon tubing is supplied at both the inlet and
outlet of the terminator. The inlet tubing is provided with a 1/16"
plastic tubing fitting for connecting to the flow cell. The inlet
tubing can be removed from the terminator and the flow cell
tubing can be connected directly. Either the outlet tubing can be
directed to the waste container, or the tubing removed, and the
terminator itself dropped to the bottom of the waste bottle.
The terminator pressure setting is not adjustable. If the
terminator fails or becomes plugged, replace the existing
cartridge with a new 40 psi replacement supplied in Kit
03-919239-90.
Installing and adjusting the Xenon lamp
The final part of installation is to install and align the xenon lamp.
This is the same procedure that you will have to do when you
replace a xenon lamp. Refer to procedures in Replacing the
Xenon Lamp beginning on page 49.