24
03-914875-00:1
To change the Unit ID, Press Enter. The screen below appears.
Unit ID (0-9)
0
At this point, you can enter any number from 0 through 9 with 0
being the default. The values 0-3 are recognized by the Star
Workstation and translates to Device Identifiers 90-93.
Turn off the power to the system and then turn the power back
on. The serial communication address is now configured.
Setting Up the Instrument Conditions
Param
Set
Call up the instrument condition setup screen by pressing the
Param Set key. The instrument parameters include:
•
TIME CONSTANT
•
PMT VOLTAGE (photomultiplier voltage)
•
Em BANDWIDTH (emission bandwidth)
•
USE TIME PROGRAM (selection of use/non-use of time
program)
•
LAMP MODE (selection of lamp turn-on/off)
Specify a time constant
TIME CONSTANT
(0.1s=1, 0.5s=2, 2.0s=3,
3
4.0s=4, 8.0s=5))
The current time constant value is displayed; the default value is
3. Time constant value ranging from 1 to 5 can be entered using
the Numeric keys. Select the desired value and press Enter.
The photomultiplier voltage setup screen will be presented. (If
the indicated time constant is acceptable, simply press the
▼
key
to go to the photomultiplier voltage setup screen, or press the
▼
key to go to the monitor screen.