Operation
ProStar 363 Fluorescence Detector
25
As an alternative, press the Escape key to return to the monitor
screen without changing the current setting.
Specify a Photomultiplier (PMT) voltage
PMT VOLTAGE (HIGH=1, MED=2, LOW=3)
1
The current PMT output range value is displayed; the default
value is 2. An output range value of 1 to 3 can be entered using
the Numeric keys.
Select the desired value and press Enter.
The Em bandwidth setting screen will be presented.
If the indicated PMT voltage is acceptable, simply press the
▼
key to go to the Em bandwidth setting screen, or press the
▲
key to go to the time constant screen. Press the Escape key to
return to the monitor screen without changing the current setting.
Specify an emission bandwidth
Em BANDWIDTH (STANDARD=1, WIDE=2)
1
The current bandwidth setting of the emission monochromator is
displayed; the default value is 1.
Input Value
Meaning
1
Emission monochromator bandwidth is
standard (15 nm).
2
Emission monochromator bandwidth is
wide (30 nm).
Select the desired value and press Enter. The time program
screen will appear.
If the indicated setting is acceptable, simply press the
▼
key to
go to the time program screen, or press the
▲
key to go to the
photomultiplier voltage setup screen. Press the Escape key to
return to the monitor screen without changing the current setting.