28
03-914875-00:1
to return to the excitation wavelength screen. As an alternative,
press the Escape key to return to the monitor screen without
changing the current setting.
To Adjust the Auto Zero
Auto
Zero
Press the Auto Zero key to adjust the FL intensity to zero.
This will affect the reading on the monitor screen and the output
signal level.
TIME EX EM FL PROG LAMP [I]
0.0
250
350
xxxx.x 1 Xe (D1)
To correct for baseline drift or if an adjustment to the zero setting
is desired, press the Auto Zero key to reset the output value to
zero.
To Place a Marker on the Recorder
Marker
Press the Marker key to indicate a marker line on the recorder.
When the Marker key is pressed, a marker of approximately 0.6
mV is output on the recorder terminal signal. This marker can be
used to indicate the time when the sample is injected or when a
change in a present condition is made during measurement.