Operation
ProStar 363 Fluorescence Detector
29
Time Program
Time
Program
Press the Time Program key, select/setup screen.
Select a time program
A time program is used when the excitation and/or emission
wavelength should be changed during the run. After the
wavelength is changed, the baseline can be autozeroed or held
as desired.
Up to 9 programs can be stored and up to 100 steps can be
included in the time programs (e.g. one program can use 100
steps, or 9 programs can use 11 steps, etc.).
TIME PROGRAM NO. (1-9)
1
The currently selected program number is displayed; the default
value is 1. The acceptable range is from 1 to 9. The time
programs are held in memory by battery backup when power to
the system is turned off.
Numeric key
→
Enter key
Enter the number for the program to be created or modified. The
relevant execution program number is changed at the same
time.
To return to the monitor screen without changing the value,
press the Escape key.