36
03-914875-00:1
Specify an excitation scan range.
Indicate the starting wavelength (minimum value). (The
information in this section is not relevant in Em Scan Mode.)
Ex SCANNING WL RANGE (200-800) nm
200 - 550
1 nm step
Enter the desired wavelength with the numeric keypad and press
the Enter key to change the value. The currently specified value
is displayed; the default value is 200. The allowable input range
is from 200 to 800 nm.
The
and
key allow you to move back and forth between
starting and ending wavelength.
Indicate the ending wavelength (maximum value)
Ex SCANNING WL RANGE (250-550) nm
200 – 550
1 nm step
Enter the desired wavelength and press the Enter key to change
the value. The currently specified value is displayed. (The default
value is 550.) The allowable input range is “start wave
50” to “start wave 350”.
Specify a fluorescence (also called emission) scan range; the
information in this section is not relevant in Ex Scan Mode.
Indicate the starting wavelength (minimum value)
Em SCANNING WL RANGE (250-850) nm
250 - 600
1 nm step
Enter the desired wavelength and press the Enter key to change
the value. The currently specified value is displayed; the default
value is 250. The allowable input range is from 250 to 850 nm.
Use the
or
keys if the present value is acceptable.
The
key returns to the monitor screen, while the
key
accesses the Em SCANNING WL RANGE end wavelength field.