Operation
ProStar 363 Fluorescence Detector
37
Indicate the ending wavelength (maximum value)
Em SCANNING WL RANGE (300-600) nm
250 - 600
1 nm step
Enter the desired wavelength and press the Enter key to change
the value. The currently specified value is displayed; the default
value is 600. The allowable input range is “start wave
50” to “start wave 350”.
The
and
key allow you to move back and forth between
starting and ending wavelength.
Note that the maximum wavelength scan range (= end
wavelength - start wavelength) for both excitation and emission
is 350 nm.
For spectral acquisition in a range beyond 350 nm, two separate
scans will have to be made.
Collecting the Spectra
Acquire
Spectrum
When the Acquire Spectrum key is pressed, the appropriate
Monochromator is scanned over the range which was set in the
section, Setting Up Spectrum Acquisition Parameters and Output
and the spectrum is stored using the indicated memory number.
When collecting a spectrum, there should be a sample in the cell
and the flow rate should be zero (0).
Specify a storage file.
SPECTRUM MEMORY NO. (1-8) MODE: Em
No.
Numeric key
→
Enter key