38
03-914875-00:1
The scanning starts with the display below.
Up to eight spectra can be stored. The currently selected scan
mode is displayed by the MODE field. The spectral memory
numbers (input values) are used for identification of stored
spectra. When a spectral memory number is specified, the
spectrum in that memory location is deleted and the new
spectrum data is stored.
While the unit is scanning, the monitor screen will present:
SCANNING SPECTRUM NO. x
Ex:xxx Em:xxx
The specified spectrum number and the current wavelength are
displayed. After the spectrum has been stored, the status
screen is presented. The stored spectral data is not retained
when power is turned OFF.
NOTE:
Stored spectral data is not retained when the power is turned
OFF.
To cancel spectral data acquisition, press the Escape key.
NOTE:
The spectrum will not be saved if the Escape key is pressed.
Setting the Recorder and Spectral Output Parameters
Output
Param
Access the recorder/spectral parameter factor setup screen by
pressing Output Param. The recorder/spectral output parameter
setup screen will then appear. This screen is used to set:
•
The recorder full scale range
•
The recorder output speed
•
The output data