Operation
ProStar 363 Fluorescence Detector
39
The Output Param key is used to access several screens which
are used to indicate information about the recorder and output of
the spectrum.
RECORD: RCD RANGE=1, RCD SPEED=2,
<1-3> 1 SPECT RCD=3
Numeric key
→
Enter key
The currently specified recorder full scale range is displayed; the
default value is 1.
Input Value
Meaning
1
RCD RANGE (Recorder range)
Enter 1 to set recorder output full scale range.
2
RCD SPEED (Recorder speed)
Enter 2 to set recorder output speed.
3
SPECT RCD (Spectral recorder)
Enter 3 to set output data format.
RECORDER RANGE (1-1000)
1000
Numeric key
→
Enter key
The recorder range only affects the recorder output, not the
processor output. The default value of 1000 means F.L.
(Fluorescence units) will be full scale (10 mV) on the
recorder output. If 100 was entered, 100 F.L. would be 10
mV. This is a means to expand the Fluorescence range to
match the recorder range.
RCD SPEED (40 nm/min=1, 60 nm/min=2)
2
Indicate the output speed for the recorder.
Numeric key
→
ENTER key
The currently specified recorder speed is displayed; the
default Recorder Speed value is 2.