40
03-914875-00:1
Input Value
Meaning
1 40
nm/min=1
2 60
nm/min=2
RECORD SPECTRUM NO. (1-8)
1
Assign a spectral number to output data
Numeric key
→
Enter key
The currently specified value is displayed; the default value is 1.
A value ranging from 1 to 8 can be entered using the Numeric
keys.
BACKGROUND SPECTRUM NO. (0-8)
0
Specify background data. The currently specified background is
displayed; the default value is 0. A value ranging from 0 to 8 can
be entered using the Numeric keys. The background spectrum
is used to compensate for fluorescence due to the solvent and/or
the flow cell. To ensure that an appropriately corrected spectrum
is used in analytical work, a new background spectrum should be
collected whenever the solvent or wavelength range is changed.
Input Value
Meaning
0
The spectrum selected by RECORD SPECTRUM NO.
is presented. No background correction is made.
1 to 8
The specified background is subtracted from the
spectrum selected by RECORD SPECTRUM NO., and
the resultant spectral data is presented.
Numeric key
→
Enter key
After the parameters are entered, the monitor screen presents;
RCD NO.
SPEED WL-RANGE MODE: Em
x 60 200-550
PRESS "START"