Example of Time Program Operation
This selection describes how a time-based program operates.
The events indicated in the figure below will be used in the
discussion.
Setting up a Time Program
c
In this example, a new time program for the activity
described in the above figure will be generated.
1. At the start of the separation, the Ex monochromator is
set to 250 nm, the Em monochromator is set to 350 nm
and the autozero function is used to define the baseline.
2. After 10 minutes, the Ex monochromator is set to 270
nm and the Em monochromator is set to 370 nm. When
the monochromators are changed, the baseline is
maintained (held).
3. After 20 minutes, the measurement of fluorescence is
completed.
Auto zero
Hold
Start
Stop
10.0
20.0
On
250
350
270
370
Time
(minutes)
Ex (nm)
Em (nm)
A/Z
Hold