Operation
ProStar 363 Fluorescence Detector
45
NOTE 1:
In time program setting, the initial 'TIME' for the first step must be 0.0.
This initial value cannot be changed or deleted.
Be sure to specify the end time of measurement in the last step. The
time program will be run until the end-of-measurement time specified at
the last step is reached.
NOTE 2:
In time program setting, provide an interval of at least 0.3 minute
between steps.
NOTE 3:
When a time program is selected, the display presents the operator with
the opportunity to indicate if the present program should be modified or if
a new program is to be entered.
If the 'MODIFY' setting is selected, the first line of the program is
displayed for editing as described below.
If the 'NEW' setting is selected for a time program, the previous time
program is deleted. The first step in the method will include the initial
values indicated in the table below. This table also includes the
allowable range for each parameter.
Initial Value
Input Range
TIME
0.0
0.0 to 600.0
Ex
200 to 850, 0
Em
250 to 900, 0
BASE 1
1,
2
START
STOP
key
TIME PROGRAM NO. (1-9)
1
1 key
→
Enter key
MODIFY=0, NEW=1
0
1 key
→
Enter key