To execute the Time Program press the Start/Stop key.
TIME
Ex Em FL PROG LAMP [I]
0.0
250 350 0.0
1
Xe
(90)
On completion of the time program, measurement is
automatically terminated and the display returns to the status
screen. At this point, the wavelength settings are those that
were on the status screen before the time program was
executed.
NOTE:
It is advisable to set the wavelengths on the monitor screen to the
same wavelengths as indicated in the Time = 0.0 of the time program.
Shutting Down the Detector
To turn the detector off, turn off the Power switch.
The set measurement conditions are retained by the memory
when the power is turned off. When the power is turned on
again, the conditions that were in place when the power was
turned off will be restored.
Flushing the Flow Cell After Use
Be sure to flush the flow cell with distilled water after using any
buffers or salts in the mobile phase. Failure to remove buffer or
salt solutions may lead to corrosion and or the formation of salt
deposits in the system.
Elapsed time
Measurement wavelength
Measured value
Program number