Maintenance and Troubleshooting
Replacing the Xenon Lamp
The xenon lamp is too hot to touch immediately after the
instrument is turned off. When replacing the xenon lamp, wait
until the old lamp cools down completely as described below.
To Cool Down the Xenon Lamp:
Press the following keys:
Param Set key
→
Enter key
→
Enter key
→
Enter key
→
Enter key
The following screen will appear.
LAMP MODE (OFF=0, ON=1)
1
On this screen, press the 0 key, followed by the Enter key to turn
off the xenon lamp. The fan will continue to run to cool down the
xenon lamp. Leave the power to the instrument on and wait for
approximately 30 minutes.
C A U T IO N
The xenon lamp must be removed from the detector prior
to shipping the unit. Failure to follow this procedure will
result in a broken lamp.