Maintenance and Troubleshooting
ProStar 363 Fluorescence Detector
53
Installing the Xenon Lamp
WA R N IN G :
SH O C K H A ZA R D
•
Potentially Dangerous Voltages are Present
within the Instrument.
•
While the POWER switch is on, a voltage of
approximately 90V is applied to the anode of
xenon lamp. When the lamp goes on, a high
voltage of approximately 30 kV is applied to
it.
•
To ensure safety, however, turn off the
POWER switch and unplug the power cord
from the outlet.
•
Before removing the xenon lamp for
replacement, turn power off and wait for at
least 30 minutes. Then, open the light
source cover.
1. Make
sure
the
Power switch is turned off before
installing the xenon lamp. To remove the light source
cover, first loosen the four mounting screws indicated in
Figure 12.