Maintenance and Troubleshooting
ProStar 363 Fluorescence Detector
55
quartz protrusion, side the wire so it is properly
positioned.
3. Attach the cathode lead plate to the cathode threaded
part, and secure the lead plate with the cathode nut.
WA R N IN G :
E XP LO SIO N H A ZA R D
•
When tightening the lamp clamp nut, hold
the metal parts on the anode and cathode
sides with fingers. Never hold the lamp
bulb when securing it. Take care not to
apply excessive force or shock to the
lamp bulb.
•
If the lamp bulb is contaminated with dust
or fingerprints, wipe it using the
furnished cleaning paper or
gauze/absorbent cotton cloth moistened
with high-quality alcohol. In this
cleaning, be careful not to apply strong
impact or shock to the lamp.
•
Be sure to mount the lamp in the
specified direction. If the mounting
direction (polarity) is wrong, the cathode
will be consumed significantly to disable
turn-on of the lamp.
Mount the lamp so that the '+' (anode)
mark on it will be positioned at the
support metal of the lamp holder.
•
If a lamp with its cathode consumed
excessively is used continuously, the
pressure inside the lamp bulb may
become too high, causing explosion. To
prevent this, replace the lamp with a new
one immediately if its cathode has been
consumed significantly