Maintenance and Troubleshooting
ProStar 363 Fluorescence Detector
63
Wave Length
Ex
Em
(200-850, 0)
nm
250 350
3 5 0
→
Enter
→
Enter
Ex
Em
MONITOR(Ex) LAMP
[I]
350
350
xxxx Xe (90)
Param Set
→
Enter
PMT VOLTAGE (HIGH=1, MED=2, LOW=3)
2
2
→
Enter
→
Escape key
Ex
Em
MONITOR(Ex) LAMP [I]
350
350
xxxx Xe (90)
4. Perform the fine adjustment as described below.
Turn the FOCUS, HORIZONTAL ADJ and VERTICAL ADJ
screws gradually in a repetitive manner while observing the
MONITOR value on the screen. Turn each screw on an
iterative basis until the MONITOR value is maximized. At
least three cycles should be used. When a maximum value
is reached, fine adjustment of the lamp position is
completed.
5. After lamp alignment is complete change display mode back
to Fluorescence.
Utility/Setup
→
3 (Display Mode)
→
Enter 1 (FL)
→
Enter
Summary of Contents for ProStar 363
Page 2: ......
Page 6: ...iv 03 914875 00 1...
Page 20: ...14 03 914875 00 1...
Page 82: ...76 03 914875 00 1...