Maintenance and Troubleshooting
ProStar 363 Fluorescence Detector
67
Cleaning of Flow Cell
1. Clean the flow cell in running water.
If the contamination cannot be removed with running water, clean
it with an ultrasonic cleaner. Be sure to put the flow cell in a
paper or plastic container and then place it in the ultrasonic
cleaner to avoid scratching the cell.
2. Wipe the surface of the flow cell with gauze moistened with
acetone or ethanol.
C A U T IO N
Be careful not to scratch the flow cell surface when
disassembling it.
Be sure to put the flow cell in a paper or plastic container
when cleaning it. Using a glass container may scratch the
cell or prevent obtaining the desired performance from the
instrument.
Figure 22 Disassembly of Flow Cell Unit
Retaining
screw 1
Cell window
Cell clamp
Flow cell
Retaining screw 2