72
03-914875-00:1
Table 4 Troubleshooting
Symptom Cause
Check Remedy
. Self-diagnosis
(initialization) is not
performed when the
Power switch is turned
on.
The power cord plug is
not securely plugged in.
Visual check
Plug in the power
cord securely.
The fuse is blown.
Check the fuse for
continuity.
Replace the fuse
with a new one.
The lamp cover is not
properly attached. (The
cover activates a safety
interlock switch.)
Visual check
Retighten the
retaining screws of
the lamp cover.
2. The keyboard does
not work.
The instrument is not
ready for operation.
Check the monitor
screen.
Press the Escape
key. The READY
status will be
presented.
3. The Auto Zero or
Acquire/Spectrum
key
does not work.
On the monitor screen,
the data mode is not set
to FL.
Check the monitor
screen.
Using Utility/Setup
Mode, setup Display
Mode to FL .
4. The Time Program
does not start.
The instrument is not
ready to operate.
Check the monitor
screen.
Press the Escape
key. The READY
status will be
presented.
5. Excessive noise
The flow cell is
contaminated.
Visual check
Clean the flow cell.
The mobile phase is not
sufficiently degassed.
A spike occurs.
Degas the mobile
phase thoroughly.
The mobile phase
contains impurities.
Check the mobile
phase.
Purify the mobile
phase, or replace it
with mobile phase of
a higher quality.
The gain level of the
recorder is too high.
Check the gain level of
the recorder.
Adjust the gain level
of the recorder.
The pump is faulty.
Check to ensure that
the operation of the
pump is smooth.
Check the pump.
Check for noise when
the pump is stopped.