Maintenance and Troubleshooting
ProStar 363 Fluorescence Detector
73
Symptom Cause
Check Remedy
The xenon lamp has
reached the end of its
useful life.
The S/N is less than
120/1.
Replace the xenon
lamp with a new
one.
6. Excessive drift
The warm-up period is
not sufficient.
—
Wait until the
instrument becomes
stable (at least ten
minutes).
Mobile phase is leaking
from a fitting.
Check each fitting.
Retighten the fitting.
An impurity (a strongly
retained compound) is
eluting from the column.
Stop feeding liquid from
the pump.
Wait until elution is
completed, or
replace the column
with a new one.
There is a significant
change in the ambient
temperature.
— Control
the
temperature
variation.
The mobile phase is not
sufficiently pure.
Check the mobile
phase solvent.
Purify the mobile
phase, or replace it
with one of a higher
quality.
7. Poor S/N ratio
Flow cell contaminated
Visual check
Clean the flow cell.
Xenon lamp off-position Check lamp energy
level
Adjust lamp
position.
Error in flow cell
mounting
Visual check
Correctly mount flow
cell.
Flow cell surface
scratched
Visual check
Replace flow cell.