74
03-914875-00:1
Error Messages
Message Description
Remedy
A/Z OVER RANGE
PRESS CLEAR KEY
TO CLEAR MESSAGE
The allowable auto zero range is
exceeded.
Change the wavelength setting.
Use a mobile phase with a lower
level of fluorescing impurities.
Clean the flow cell.
LAMP ERROR
The lamp goes off during analysis,
the lamp does not turn on, or a
momentary power interruption is
encountered.
Press the Clear key to reset the
system after a momentary power
interruption.
If the lamp does not start, turn the
power off and turn it on again.
If the lamp is at the end of its
useful life time, replace it with a
new one.
Ex(Em) SIDE OF WL
DRIVE MECHANISM
ERROR
The monochromator cannot
properly set during the self-
diagnosis test (initialization).
Turn power on again.Notify the
nearest Varian service center.
ROM ERROR
Error in ROM
Notify the nearest Varian service
center.
C-RAM ERROR
Error in RAM
Notify the nearest Varian service
center.
CHECK TIME
PROGRAM
The allowable maximum number
of time program steps is
exceeded.
Remove unnecessary steps from
the time program.
OVER HEAT! LAMP
WAS TURNED OFF
The fan has stopped and the
temperature sensor was activated.
The fan is faulty. Notify the
nearest Varian service center.
The lamp becomes too hot. It is
not cooled down by the fan after
turn-off.
After the lamp cools down, turn on
power on again.
LAMP WAS TURNED
OFF BY LAMP COVER
OPEN
TURN OFF THE
POWER
Lamp cover is open.
Close the lamp cover.
Turn on power supply again.