Maintenance and Troubleshooting
ProStar 363 Fluorescence Detector
75
Message Description
Remedy
NO SPECTRA OR
SPECTRA ARE
MISMATCHED
PRESS CLEAR KEY
TO CLEAR MESSAGE
No spectral data, scan mode in-
correct, fixed wavelength incorrect,
background WL range doesn’t
include entire spectrum WL range,
or background slit position differs
from spectrum slit position.
Select correct parameters.
PARAMETER ERROR Error in parameter data (data
cannot be retained due to
EEPROM error
Turn on power again.
Contact your nearest Varian
service center.
NOTICE:
An input value stored in this instrument may be deleted if certain
error messages are presented.
If this is the case, enter the relevant value again.
In the event of PARAMETER ERROR, the wavelength
calibration value is deleted.
In this case, it is necessary to perform the wavelength calibration
procedure.