78
03-914875-00:1
Marker
Marker output of approx. 0.6 mV at recorder
terminal (corresponding to about 6 graduations
on chart)
Instrument parameter setting
Following parameters are settable and stored
in battery-backed-up memory when power is
turned off.
Full scale range
Spectrum measuring range
Photomultiplier applied voltage
Offset value
Time program
# of programs
Settable time
No. of steps
storable
Programmable
items
9
up to 600 minutes in
increments of 0.1 minute
Up to 100 steps for a total
of 9 files
Excitation wavelength,
emission wavelength,
baseline processing
Display function
40 characters
×
2 lines on backlighted LCD
Communication function
RS-422/485 serial communication
External output terminals
Analog output
For processor : 1 V full scale
For recorder
: 10 mV full scale
External I/O contact terminals
Time program start, error input/output, busy
output. Auto zero input, marker input, lamp off
input
Flow cell capacity
Standard cell 12
µ
L (irradiated capacity)
Flow cell withstand pressure
29 ATM (425 psi)
Operating temperature range
4 to 35
°
C
Operating humidity range
45 to 80% RH (non-condensing)
Power requirement
100-115, 220-240 V AC, 50/60 Hz
Power consumption 360 VA
Dimensions 292 mm W x 470 mm D
×
305 mm H
Weight
Approx. 19 kg