2000 MS Hardware Maintenance
102
Mass Spectrometer Maintenance
A Heating
Cable
H O-ring
B
Boot
I
Transfer Line Tip
C
Nut
J
Heating Cable Slot
D Ferrule
K Nose
Clip
E Tool
L Bayonet
Mount
F
Nose
M
Analyzer Assembly Tongue
G
Nose Hole
N
Analyzer Assembly Lock-Down
Tabs
Transfer Line Assembly
Remove the Analyzer Assembly
Retract transfer line before removing analyzer assembly.
NOTE: Be sure the transfer line is retracted. Otherwise, you will not be able to
remove the analyzer assembly without damaging the analyzer.
To remove the analyzer assembly, proceed as follows:
1. Remove the top cover of the Saturn GC/MS by grasping both sides and lifting
up.
2. Unplug the trap heater harness located near the top of the instrument.
3. On the side of the analyzer assembly (near the transfer line), push out the
locking tabs on the power ribbon cable. This releases the cable.
4. Pull the ribbon cable out and move it away from the analyzer.
5. Push down and spread the two analyzer release tabs.