03-914978-00:1
103
Mass Spectrometer Maintenance
NOTE: Some Saturn systems have a transfer line removal flap warning that will
block the locking tabs. If such a flap is present, tip it out of the way during the
procedure and return it to its original position once the analyzer is replaced.
6. Tilt the rear end up carefully to remove the analyzer.
7. Move the analyzer assembly toward the rear to free the front tab.
8. Place the analyzer upside down on a flat surface.
NOTE: To prevent contamination when touching parts of the trap or the electron
multiplier, wear gloves.