03-914978-00:1
11
Functional Description
A Main Power Switch (rear
panel)
J
Cal Gas Vial
B Service Switch
K Vent Valve
C Transfer Line Heater
L
RF Coil
D Trap Heater
M RF Coil Adjustment Screw
E Manifold Heater
N Transfer Line
F LED
O Turbomolecular
Pump
G Pneumatics Manifold
P Cooling Fans
H CI Cal Gas Adjust
Q CI Shutoff Valve
I
Cal Gas Adjust
Saturn 2000 GC/MS Mass Spectrometer
Cooling Fans
To prevent overheating, do not block air intakes.
Turbomolecular Pump System
Two fans mounted on the rear of the spectrometer cool the unit. The analyzer
compartment fan draws air from the back, blowing it directly on the bearing end
of the turbomolecular pump in the analyzer compartment. The air then flows past
the manifold electronics and out the front of the instrument. The turbomolecular
pump controller supplies power to the analyzer compartment fan.