6.16 CPNOESYPS—Cross-Polarization Phase-Sensitive NOESY
01-999162-00 C0402
VNMR 6.1C User Guide: Solid-State NMR
104
Parameters
cpcosyps
uses
UNITY
INOVA and UNITYplus parameters
tpwr
,
tpwrm
,
dpwr
,
cppwr
,
dipolr
,
crossp
,
dblvl2
,
pw
, and
cntct
. See
for a description of these
parameters.
xpol
is set to
'n'
for preparation by direct polarization;
xpol
is set to
'y'
for
preparation by cross-polarization.
pw
is the observe 90
°
pulse (
xpol='n'
) or the proton 90
°
pulse (
xpol ='y'
).
pw
is in
microseconds.
pwx
is the mixing pulse, in microseconds. If
xpol='n'
,
pwx=pw
; if
xpol='y'
, it is set
to the observe 90
°
pulse.
srate
is the measured MAS spinning speed. CPCOSYPS does not require rotor speed
control. However, with rotor speed control, set
sw1
equal to
srate
to remove spinning
sidebands in both dimensions.
phase=1,2
for the hypercomplex method (use
wft2da
for the 2D FT);
phase=3
for
TPPI (use
wft2d(1,0,0,0)
for the 2D FT).
Reference
Wu, G.; Wasylishen, R. E. Organometallics 1992, 11, 3242.
6.16 CPNOESYPS—Cross-Polarization Phase-Sensitive
NOESY
The CPNOESYPS sequence is a 3-pulse 2D exchange correlation sequence similar to the
high-resolution NOESY experiment. CPNOESYPS can be run as a direct polarization
experiment or the first 90
°
pulse can be substituted with a cross-polarization pulse element.
is a diagram of this sequence.
Figure 44. CPCOSYPS Pulse Sequence
pw
cntct
cntct
(tpwrm)
(crossp)
(dipolr)
Tx
Dec
at
d1
A
C
pwx
d2