6.20 REDOR1—Rotational Echo Double Resonance
01-999162-00 C0402
VNMR 6.1C User Guide: Solid-State NMR
112
Applicability
REDOR1 is available only on
UNITY
INOVA and UNITYplus systems.
Macro
The macro
redor1
converts a parameter set, obtained with XPOLAR or XPOLAR1, for
the REDOR experiment with
xy8
π
pulses on the observe channel. Observe and decoupler
power levels and the proton 90
°
pulse width are retained. Third channel power levels and
pulse width must be calibrated (see
p1
below). Default parameters yield an array of spectra
for 2 to 32 rotor cycles. Alternate spectra are obtained without and with third channel
π
pulses. For best results, adjust
tof
and
dof2
(see
dof2
below) so that the peak of interest
is on-resonance.
Parameters
redor1
uses
UNITY
INOVA and UNITYplus parameters
tpwr
,
tpwrm
,
dpwr
,
cppwr
,
dipolr
,
crossp
,
dblvl2
,
pw
, and
cntct.
See
for a description of these
parameters.
rdflag
is set to
'y'
to apply
π
pulses to both the observe and third channel and thus
obtain REDOR modulation of the peak intensity.
rdflag
is set to
'n'
to apply
π
pulses
only to the observe channel and thus obtain observe peak intensity without REDOR
modulation. Typically, the array
rdflag='n','y'
is used.
Figure 49. REDOR1 Pulse Sequence
'cycles' - 1 times
pw
cntct
(crossp)
(dipolr)
cntct
2*pw
(tpwrm)
p1
p1
<if rdflag = 'y'>
Tx
Dec2
Dec
n*(1/srate)
d1
C
A