2.11 Useful Conversions
01-999162-00 C0402
VNMR 6.1C User Guide: Solid-State NMR
44
1
H
T
1
Through
13
C Cross-Polarization
l
H T
1
can be measured using the XPOLAR1 pulse sequence by setting it up to perform a
standard inversion-recovery experiment on the protons followed by cross-polarization of
the remain
1
H magnetization to the carbons.
illustrates the pulse sequence.
2.11 Useful Conversions
Convert from 90
°
pulse width to
γ
H:
Convert from field strength in gauss to field strength in gauss:
1
H:
γ
H (kHz)
≈
4.3 *
γ
H (gauss)
13
C
:
γ
H (kHz)
≈ γ
H (gauss)
Convert from rf fields to power levels:
P (watts)
∝ (γ
H)
2
Figure 15. Pulse Sequence for Measuring
l
H T
1
13
C
xpol='y'
1H
cntct
d1
p180
d2
pw
(crossp)
(dipolr)
cntct
(tpwrm)
γ
H
2
kHz
(
)
250
90
°
pulsewidth
--------------------------------------
µ
s
( )
=