TECHNICAL INFORMATION
64
87-900-854-01 (C)
•
Enter the value by pressing the PUMP CURRENT
pushbutton.
•
Press the CYCLE NUMBER pushbutton to select the
second number.
S P E E D
T H R E S H O L D
S E L E C T I O N : X X K R P M
•
Enter the value by pressing the PUMP CURRENT
push-button, and the display shows:
R U N - U P
T I M E
S E L :
X X h
X X m
X X s
where:
RUN-UP TIME
= is the interval time from
start to speed threshold value in hours, minutes,
seconds. Select from 00 to 99 hours, and from 00 to
59 minutes or seconds.
•
Select the run-up time according to the chamber
volume and/or operating cycle feature (see the
preceeding cycle diagram) by pressing the CYCLE