Check that pump rotor is free to rotate then press the
STOP RESET pushbutton twice to start the pump.
NOTE
After the starting phase if the output connection is
shorted (output current higher than 2.2 A), the display
will be as shown in the following figure.
F A U L
T
:
S H O R T
C
I
R C U
I
T
Check connections and shortages between pump and
controller, then press the STOP RESET pushbutton
twice to start the pump.
NOTE
If the pump is stopped by an emergency stop signal
provided via a remote contact, the display will be as
shown in the following figure.
S Y S T E M
O V E R R
I
D E
Remove the controller power cable and check the
emergency condition.
Then reconnect the power cable and press the START
pushbutton to start the pump.
The following message is displayed if the fuse of the
transformer secondary blows:
C O N T R O L
L E R
F A
I
L U R E
Since no operation can be carried out, contact
VARIAN.
ACCESSORIES AND SPARE PARTS
DESCRIPTION
PART NUMBER
J1 input mating connector
969-9853
P7 mating plug 969-9854
Mains cable (European plug, 3 m long)
969-9957
Mains cable (American plug, 120 V, 3 m long)
969-9958
Fan extension cable
969-9949
OPTIONS
DESCRIPTION
PART NUMBER
RS 232 kit
969-9851
P2 output mating connector
969-9852
Controller to pump extension cable (5 m extension)
969-9950