VHS-4, VHS-6 and VHS-250 Diffusion Pumps
1-12
D
R
A
FT 3/9
/
05
❑
The minimum rating of this system should be 85% of the
maximum power rating of the diffusion pump.
❑
The Quick Cool Coil feed line, located at the boiler plate,
should be controlled by a separate three-way valve (open,
closed, and vent to atmosphere).
2. Connect the quick cool drain to an open drain which is below the inlet connection of
the Quick Cool Coil to ensure that the Quick Cool Coil is completely drained when the
cooling water supply is turned off and the pump is vented to atmosphere.
1.2.6 Electrical Connections
WARNING
Diffusion pump heaters operate at voltages high enough to kill
through electrical shock.
❑
During installation, check the drawings and be sure to attach
all hazard warning and caution labels.
❑
Always break the primary circuit of the power supply when
direct access to the heater or wring is required.
Read the safety section in the front of this manual.
The diffusion pump has been designed to operate at a specific voltage. The voltage is
specified on the label that is mounted on the side of the pump.
1. Verify the heater rating by measuring the resistance of the heater circuit and comparing
it to the values in Table 1-1 on page 1-1.
2. Make the electrical connections in the junction box located near the foreline. The
electrical supply should not be more than ±5% of the rated voltage.
CAUTION
All electrical connections should be made in accordance with
all applicable state, local, and/or industrial codes.
Summary of Contents for VHS-250
Page 2: ...Copyright 2005 Vacuum Technologies VHS 4 VHS 6 and VHS 250 Diffusion Pumps ...
Page 4: ...VHS 4 VHS 6 and VHS 250 Diffusion Pumps This page intentionally left blank ...
Page 6: ...VHS 4 VHS 6 and VHS 250 Diffusion Pump This page intentionally left blank ...
Page 8: ...VHS 4 VHS 6 and VHS 250 Diffusion Pumps This page intentionally left blank ...
Page 10: ...VHS 4 VHS 6 and VHS 250 Diffusion Pumps This page intentionally left blank ...
Page 12: ...VHS 4 VHS 6 and VHS 250 Diffusion Pumps This page intentionally left blank ...
Page 34: ...VHS 4 VHS 6 and VHS 250 Diffusion Pumps This page intentionally left blank D R A F T 3 9 0 5 ...
Page 50: ...VHS 4 VHS 6 and VHS 250 Diffusion Pumps This page intentionally left blank D R A F T 3 9 0 5 ...
Page 56: ...VHS 4 VHS 6 and VHS 250 Diffusion Pumps This page intentionally left blank D R A F T 3 9 0 5 ...
Page 59: ......