Leave the introducer placed until the haemodynamic profile returns to normality. Perform su-
ture as usually.
•
Further dilatation of stented segments
If the stent size results inadequate with respect to vessel diameter, stent post-dilatation technique
may be applied with a larger balloon. To do this, a balloon catheter is inserted into the stented area
(it is not needed to be compliant) and dilatation is performed according to the manufacturer’s recom-
mendations.
As a further recommendation, a “small stent” must not be post-dilated over 3.30mm, a “medium stent”
must not be post-dilated over 4.25mm and a “large stent” must not be post-dilated over 5.25mm.
Withdrawing an unexpanded stent
If the stent does not cross the lesion, it may need to be removed without expanding. If any resistance
is not detected, check that the guiding catheter is coaxially situated with respect to the stent and
carefully withdraw the stent towards the guiding catheter. If any unusual resistance is noticed during
its withdrawal, extract the whole system as a single unit by means of fluoroscopic imaging according
to the indicated instructions in part 5.3 of this leaflet.
8. Warranty
The product and all its components are designed, manufactured, tested and packed taking all reasonable
precautions. LVD Biotech guarantees the product up to its expiry date providing the packaging is not broken,
tampered with or damaged.
Catalogue reference
Manufacturer
Consult instructions
for use
Batch no.
Sterilized with ethylene
oxide
Do not re-use
Expiry date
Limitation in storage
temperature
Do not re-sterilize
Keep away from sunlight
Keep dry
Do not use if the pack is
damaged
STERILIZE
2
STERILIZE
2
STERILIZE
2
STERILIZE
2
STERILIZE
2
STERILIZE
2
STERILIZE
2
STERILIZE
2
STERILIZE
2
STERILIZE
2
STERILIZE
2
STERILIZE
2
STANDARDISED SYMBOLS FOR USE IN LABELLING MEDICAL DEVICES
(EN 980-2008 STANDARD)
INSTRUCTIONS FOR USE
CoCr CORONARY STENT SYSTEM
architect
therapies for living
Instrucciones_Architect REV2.indd 8
14/01/15 10:35