4. Operation
EzRay Air Wall (VEX-S300W) User Manual
25
EN
G
LISH
3. When tooth types are selected, the exposure angles are automatically set
according to the tooth type. To check the default exposure angles, refer to the
following table.
Tooth Type
Angle of Inclination
Incisor
Maxilla: +40° ~ +50°
Mandible: -22° ~ -28°
Canine
Maxilla: +40° ~ +50°
Mandible: -17° ~ -23°
Molar/Premolar
Maxilla: +25° ~ +35°
Mandible: -2° ~ -8°
Bitewing
+3° ~ +12°
Since the angles of inclination for the maxilla incisor and
canine are the same, the exposure time of the canine is
applied to both the maxilla incisor and canine.
Refer to the following figure to see the angle for molar/premolar.