9. Product Specifications
56
EzRay Air Wall (VEX-S300W) User Manual
To assure line voltage quality, a separate 3-core
grounded power cable connected directly to a central
distribution panel with an over-current circuit breaker
rated for 10A must be used.
The mains resistance should not exceed 0.5
Ω.
The system is available with a fixed tube current specification
based on the user's choice.
Power Supply is specified as a part of the ME EQUIPMENT.
Power plugs may have various specifications for each
country.
9.4
Environmental Specifications
Item
Description
During operating
Temperature
10 ~ 35
℃
Relative humidity
30 ~ 75 %
Atmospheric pressure
860 ~ 1060 hPa
Transport and storage
Temperature
-10 ~ 60
℃
Relative humidity
10 ~ 75 % non-condensing
Atmospheric pressure
860 ~ 1060 hPa
Summary of Contents for EzRay Air Wall
Page 1: ...English User Manual Model VEX S300W Version 1 38 Full version...
Page 2: ......
Page 3: ......
Page 4: ......
Page 5: ......
Page 7: ...Notice vi EzRay Air Wall VEX S300W User Manual This page intentionally left blank...
Page 27: ...3 System Overview 18 EzRay Air Wall VEX S300W User Manual This page intentionally left blank...
Page 47: ...4 Operation 38 EzRay Air Wall VEX S300W User Manual This page intentionally left blank...
Page 51: ...6 User Maintenance 42 EzRay Air Wall VEX S300W User Manual This page intentionally left blank...
Page 57: ...9 Product Specifications 48 EzRay Air Wall VEX S300W User Manual...
Page 58: ...9 Product Specifications EzRay Air Wall VEX S300W User Manual 49 ENGLISH...
Page 63: ...9 Product Specifications 54 EzRay Air Wall VEX S300W User Manual 4 Tube Dimensions mm...
Page 83: ......
Page 84: ......
Page 85: ......