Appendix
60
EzRay Air Wall (VEX-S300W) User Manual
A.3 X-ray Dose Data
The X-ray dose data is extracted from the X-ray Dose Test Report for the
EzRay Air
Wall (VEX-S300W)
. The IEC collateral standards have measured the X-ray doses of
EzRay Air Wall (VEX-S300W)
in the test report. The
EzRay Air Wall (VEX-S300W)
has been designed by Part 1. General Requirements for Safety, IEC 60601-1-3.
Test Condition
Model Name
VEX-S300W
Tube Model Name
V1-650304
Generator Model Name
DG-S0102V1 (Inverter type)
Loading Factor
65 kV, 3.0 mA
1. X-ray Dose Table
Test Equipment
Instrument
Manufacturer
Model
S/N
Multi-Dose Meter
Raysafe
Unfors Xi mAs
/ Unfors Xi R/F & MAM
163288 / 161834
Dose Area Product (DAP) Table (65 kVp, 3.0 mA, SSD 200 mm)
Dose Table
FOV: Ø 6 cm
FOV : 4 x 3 cm
FOV : 2 x 3 cm
t (s)
mGy
Dose (mGy.cm
2
)
0.05
0.222
2.63
1.12
0.56
0.08
0.357
4.71
2.00
1.00
0.10
0.443
6.09
2.59
2.29
0.15
0.663
9.55
4.06
2.03
0.20
0.909
13.01
5.53
2.76
0.25
1.101
16.47
6.99
3.50
0.30
1.326
19.93
8.45
4.23
0.45
1.781
30.32
12.87
6.44
0.50
2.212
33.78
14.34
7.17
Summary of Contents for EzRay Air Wall
Page 1: ...English User Manual Model VEX S300W Version 1 38 Full version...
Page 2: ......
Page 3: ......
Page 4: ......
Page 5: ......
Page 7: ...Notice vi EzRay Air Wall VEX S300W User Manual This page intentionally left blank...
Page 27: ...3 System Overview 18 EzRay Air Wall VEX S300W User Manual This page intentionally left blank...
Page 47: ...4 Operation 38 EzRay Air Wall VEX S300W User Manual This page intentionally left blank...
Page 51: ...6 User Maintenance 42 EzRay Air Wall VEX S300W User Manual This page intentionally left blank...
Page 57: ...9 Product Specifications 48 EzRay Air Wall VEX S300W User Manual...
Page 58: ...9 Product Specifications EzRay Air Wall VEX S300W User Manual 49 ENGLISH...
Page 63: ...9 Product Specifications 54 EzRay Air Wall VEX S300W User Manual 4 Tube Dimensions mm...
Page 83: ......
Page 84: ......
Page 85: ......