Smart Plus (PHT-35LHS)
Technical Manual
VT-TECH-Smart Plus-001
168 / 188
COPYRIGHT© 2018 VATECH
.
ALL RIGHTS RESERVED
▶
Laser Motor Stop Torque Set(Initial-Value : 0100 (0~1499))
2800 =>(SPM_TSRT_0001) TEMPLE_M_RUN_TORQUE
▶
Temple Support Motor Run Torque Set(Initial-Value : 2800 (0~1499))
0100 =>(SPM_TSST_0001) TEMPLE_M_STOP_TORQUE
▶
Temple Support Motor Stop Torque Set(Initial-Value : 0100 (0~1499))
20000 =>(SPM_GMRT_0001) G_M_RUN_TORQUE
▶
3D Camera Motor Run Torque Set(Initial-Value : 20000 (0~65535))
01000 =>(SPM_GMST_0001) G_M_STOP_TORQUE
▶
3D Camera Motor Stop Torque Set(Initial-Value : 1000 (0~65535))
20000 =>(SPM_HMRT_0001) H_M_RUN_TORQUE
▶
CEPH Sensor Motor Run Torque Set(Initial-Value : 20000 (0~65535))
01000 =>(SPM_HMST_0001) H_M_STOP_TORQUE
▶
CEPH Sensor Motor Stop Torque Set(Initial-Value : 1000 (0~65535))
~~~~~~~~~~~~~~~~~~~~~~~~~~~~
00600 =>(SPM_CWFQ_0600) CWLED_PULSE_HZ
▶
CEPH White LED Frequency set(Initial-Value : 600
)
00050 =>(SPM_CWDT_0050) CWLED_PULSE_DT
▶
CEPH White LED Duty set(Initial-Value : 50)
~~~~~~~~~~~~~~~~~~~~~~~~~~~~
0000=>(spm_DROP_000x)DOOR_TYPE
0: NO door switch Mode
1: Using door switch mode
▶
Door Lock Function Setting
0000: No Use Door Lock Function
0001: Use Door Lock Function
0001=>(spm_TITY_000x)TILT_TYPE
0: NO tube tilting Mode (for only Pano Model)
1: Using tube tilting mode
▶
Tube Tilting Function Setting
0000: No Use Tube Tilting Function (PANO Only model)
0001: Use Tube Tilting Function (default)
0001=>(spm_RTOP_000x)RETURN_TYPE
0: NO auto return Mode
1: Using auto return mode
▶
After Image Acquisition, Rotating Unit Moves to Lamp Position Automatically
0000=>(spm_EPOP_000x)EXPOSURE_TYPE
0: Normal Mode
1: Sound delay mode
▶
Exposure switch function setting during “Ready to image” guide speech
0000: Override guide speech
0001: Wait for guide speech completion
0200=>(spm_FANT_xxxx)FAN_On_Temp (0400 = 40 degrees C)
▶
Temperature Setting to Operate Tube Cooling Fan Automatically (Initial-Value : 0400 (1~65534))
Ex) 400 = Tube : if the internal temperature of tube reaches to 40
℃
, the cooling fan is automatically ON
0000=>(spm_TSMD_000x)TEMPLE_SUPPORT_MOVING_DISABLE_SET
0: Normal Mode
1: Temple support disable mode
▶
Temple support moving setting
0000: Moving Enable
0001: Moving Disable
0000 =>(SPM_CBMD_000x)CT_BEAM_MOVING_DISABLE_SET
0: Normal Mode
1: CT beam moving disable mode
▶
CT beam moving setting
0000: Moving Enable
0001: Moving Disable